MP3 player headphones may hinder pacemakers

By Will Dunham

WASHINGTON (Reuters) - Headphones used with MP3 digital music players like the iPod may interfere with heart pacemakers and implantable defibrillators, U.S. researchers said on Sunday.

The MP3 players themselves posed no threat to pacemakers and defibrillators, used to normalize heart rhythm. But strong little magnets inside the headphones can foul up the devices if placed within 1.2 inches of them, the researchers told an American Heart Association meeting in New Orleans.

Dr. William Maisel of the Medical Device Safety Institute at Beth Israel Medical Center in Boston led a team that tested eight models of MP3 player headphones, including clip-on and earbud types, in 60 defibrillator and pacemaker patients.

They placed the headphones on the patients’ chests, directly over the devices. The headphones interfered with the heart devices in about a quarter of the patients — 14 of the 60 — and interference was twice as likely in those with a defibrillator than with a pacemaker.

Another study presented at the meeting showed that cellular phones equipped with wireless technology known as Bluetooth are unlikely to interfere with pacemakers.

A pacemaker sends electrical impulses to the heart to speed up or slow cardiac rhythm. The magnet, however, could make it deliver a signal no matter what the heart rate is, possibly leading to palpitations or arrhythmia, the researchers said.

An implantable cardioverter defibrillator signals the heart to normalize its rhythm if it gets too fast or slow. A magnet could de-activate it, making it ignore an abnormal heart rhythm instead of delivering an electrical shock to normalize it.

The devices usually go back to working the right way after the headphones are removed, the researchers said.

“The main message here is: it’s fine for patients to use their headphones normally, meaning they can listen to music and keep the headphones in their ears. But what they should not do is put the headphones near their device,” Maisel said in a telephone interview.

So that means people with pacemakers or defibrillators should not place the headphones in a shirt pocket or coat pocket near the chest when they are not being used, drape them over their chest or have others who are wearing headphones rest their head on the patient’s chest, Maisel said.

Most of the headphones had magnetic field strengths more than 20 times higher than the threshold for interfering with pacemakers or defibrillators, he said. They were made by Sony Corp, Philips Electronics and others.

MP3 players like Apple Inc’s iPod are popular consumer electronic devices. In January, a U.S. Food and Drug Administration researcher said the iPod is unlikely to interfere with pacemakers because it does not produce enough of an electromagnetic field to interfere with the devices.

Brian Markwalter of the Consumer Electronics Association industry group urged consumers to inform themselves about proper use of products with magnets, and encouraged people with pacemakers to understand how headphones can be used safely.

(Editing by Maggie Fox)

Source

--------------------------------------------------------------------------------------------
Related Posts:


By Michael Kahn LONDON (Reuters) - Pacemakers and defibrillators of the future may generate an extra power boost from a surprising energy source: The heart itself. Using a microgenerator powered by heartbeats, a British team said on Monday their experiment produced nearly 17 percent of the electricity needed to run an artificial pacemaker. This means the next era

Full Post: Heartbeats may power future pacemakers
--------------------------------------------------------------------------------------------

NEW YORK (Reuters Health) - Patients with heart rhythm disturbances who have an implantable heart defibrillator are particularly vulnerable to air pollution, a Swedish study indicates. In patients with these devices, known as implantable cardioverter defibrillators, or ICDs, exposure to air pollution may rapidly (within 2 hours) prompt ventricular arrhythmia — a potentially life-threatening condition in

Full Post: Air pollution may prompt abnormal heart rhythm
--------------------------------------------------------------------------------------------

By Julie Steenhuysen CHICAGO (Reuters) - Implantable heart defibrillators — which have been found to save lives in younger patients — appear to have the same benefit for people over 70, U.S. researchers said on Tuesday. The study suggests the devices, known as implantable cardioverter defibrillators or ICDs, should be offered to people in their 70s despite

Full Post: Implantable defibrillators save older patients too
--------------------------------------------------------------------------------------------

WASHINGTON (Reuters) - U.S. doctors have found the Bee Gees 1977 disco anthem “Stayin’ Alive” provides an ideal beat to follow while performing chest compressions as part of CPR on a heart attack victim. The American Heart Association calls for chest compressions to be given at a rate of 100 per minute in cardiopulmonary resuscitation (CPR).

Full Post: Disco tune “Stayin’ Alive” could save your life
--------------------------------------------------------------------------------------------

NEW YORK (Reuters Health) - Injuries to the head and face are fairly common in women’s field hockey, raising the question of whether players need more protective gear, according to researchers. In a study of six college field hockey teams, researchers found that of 253 athletes followed for two seasons, 22.5 percent suffered an injury to

Full Post: Head injuries common in women’s field hockey